Introduction
Portable Multi-parameter Patient Monitor CMS8000 can measure ECG, RESP, NIBP, SpO2, and dual-channel TEMP.(Standard)
-12.1'' TFT color LCD, multi-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakh, Polish, Czech).
-Software updating by USB interface
-Optional:IBP,ETCO2,12-lead ECG,thermal printer,VGA output,trolley
-All-round monitor for adult, pediatric and neonate.
ECG Heart rate(HR)
ECG waveform
Arrhythmia and ST-segment analysisRESPRespiration rate(RR)
Respiration waveformSpO2 Oxygen saturation(SpO2)
Plethysmogram(PLETH) waveform
Pulse rate(PR)
Bar graphNIBPSystolic pressure(SYS), Diastolic pressure(DIA), Mean pressure(MEAN)TEMP T1, T2, TDIBP(optional) SYS,DIA,MAP
CH2:SYS,DIA,MAP
IBP waveformCO2(optional) EtCO2
InsCO2:Inspired Minimum CO2
AwRR:Air Way Respiration
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc.
ECGLead mode: 3-lead or 5-lead
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: 5-lead, dual-channel
3-lead, single-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/sHRMeasurement and alarm range: 15~350 bpm
Accuracy: ±1 % or ±1 bpm, whichever is greater
Alarm accuracy: ± 2 bpm
Resolution: 1 bpm
ST-segment monitoring:
Measurement and alarm range: -2.0 mV ~ +2.0 mV
Accuracy: -0.8 mV~+0.8 mV ±0.04 mV or ±10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,
PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Pacemaker: yesRESPMethod: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150 rpm
Resolution: 1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm: 10~40s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/sNIBPMethod: Oscillometry
Mode: Manual/AUTO/STAT
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult mode: 297±3 mmHg
Pediatric mode: 240±3 mmHg
Neonatal mode: 147±3 mmHgSpO2Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%: unspecified PRMeasurement and alarm range: 30 ~ 250 bpm
Measurement accuracy: ±2 bpm or ±2%, whichever is greaterTEMPChannel: dual-channel
Measurement and alarm range: 0 ~ 50ºC
Resolution: 0.1 ºC
Accuracy: ±0.1 ºCEtCO2Method: Sidestream or Mainstream
Measuring Range: 0~150mmHg
Resolution:
0~69 mmHg, 0.1 mmHg
70~150 mmHg, 0.25 mmHg
Accuracy:
0~40 mm Hg ±2 mm Hg
41~70 mm Hg ±5%
71~100 mm Hg ±8%
101~150 mm Hg ±10%
AwRR Range: 2~150 rpm
AwRR Accuracy: ±1BPM
Apnea Alarm: YesIBPChannel: dual-channel
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and Alarm Range: -10~300 mmHg
Resolution 1 mmHg
Accuracy: ±2% or 1mmHg, whichever is greaterPower supply100-240V~, 50/60HzSafety classificationClass I, type CF defibrillation-proof applied part