ElectrochemicalPrinciple Medical oxygen O2 Sensor KE series for Ventilator Machine
1.Product Description:
Oxygen Sensor KE Series (KE-12, KE-25, and KE-50) are
unique galvanic cell type oxygen sensors developed in Japan
in 1985. Their most notable features are long life expectency,
excellent chemical durability, and no influence by CO2. KE Series
oxygen sensors are ideal to meet the ever-increasing demand
for oxygen monitoring in various fields such as combustion gas
monitoring, the biochemical field, medical applications, domestic
combustion appliances, etc.
Features:
Long life - 2.5~10 years in ambient air
Virtually no influence from CO2, CO, H2S, NOx, H2
Low cost
Operates in normal ambient temperatures
Stable output signal
No external power supply required for sensor operation
No warmup time is required
2.Main Application:
Applications:
Medical - Anesthetic instruments, respirators,
oxygen-enrichers
Biotechnology - Oxygen incubators
Food industry - Refrigeration, greenhouses
Safety - Air conditioners, oxygen detectors, fire
detectors
3.Main Features:
* Long life - 2.5~10 years in ambient air
* Virtually no influence from CO 2 , CO, H 2 S, NOx, H 2
* Low cost
* Operates in normal ambient temperatures
* Stable output signal
* No external power supply required for sensor operation
* No warmup time is required
4.Specifications:
Item ModelKE-12KE-25KE-50Measurement range0~100% O2Accuracy (Note 1)±1% full scale±1% full scale±2% full scaleOperating conditionsAtmospheric pressure811hPa~1216hPaTemperature5deg~40degCRelative humidity10-~90%RH (no condensation)CResponse time (90%) (Note 2)approx 5 sec.approx 15 sec.approx 60 secInitial output voltage under standard test conditions 19.0~35.0mV10.0~15.5mV47~65mV Atmospheric pressure1013±5hPaTemperature25 deg ±1deg CRelative humidity60±5%RHLife expectancy at 20deg C in normal air (Note 3)approx 2.5 yrsapprox 5 yrsapprox 10 yrs
Notes:
1) When calibrated at both 0% and 100% of O2,
accuracy in the range from 0-100% O2 shall be
within ±1% of full scale for both KE-12 and KE-25
and ±2% of full scale for KE-50.
2) Sensors should be used under conditions
where the air exchange is greater than
200~300ml per minute in order to obtain the
response speed as shown in the Specifications.
3) Life expectancy at 20 deg C in normal air is defined
as the period until sensor output drops to 70% of
its original value.
6. Sensitivity Characteristics: