GD6100 Automatic Precision Oil Dielectric Loss Testers is a high-precision instrument of testing dielectric loss angle of insulating oil and other liquid, integrated with main components of cup oil of dielectric loss, temperature controller, temperature sensor, dielectric loss test power Bridge, AC test power, standard capacitors.
Features
1. Advanced high-frequency induction heating mode, heat evenly, fast speed, easy to control.
2. All digital technology, automated measurement.
3. Large LCD display, full English menu, the results can be printed.
Technical Specifications
Conditions-5ºC-+40ºC, RH<80%Power SupplyAC 220V +/-10%High output1000V ~ 2200V every 200VAccuracy2%Capacity50VA Output Current 25mATemperature induction furnaceMaximum Power 1500WTemperature induction furnaceMaximum Power 1500WTemperature control range<100degreeTemperature measurement error+/-0.5degreeTemperature error+/-1degreeTemperature timeLess than 20mins, at room temperature to 90 degreeMeasuring rangeTgδ Unlimited, Cx 15PF ~ 300PFAccuracyTgδ: +/- (1.0% of the reading + 0.020%)Cx+/-(1.0% of the reading+ 0.5PF)Dimensions450 (L)*310 (W)*360 (H)Weight18kg